Combined Biopsy and Imaging-Guided Microwave Ablation by Using a Coaxial Guiding Needle

Authors:

Yi-Wei Wu ,

Tan Tock Seng Hospital, SG
X close

Gabriel Chan,

Tan Tock Seng Hospital, SG
X close

Ivan Kuang Hsin Huang,

Tan Tock Seng Hospital, SG
X close

Justin Kwan,

Tan Tock Seng Hospital, SG
X close

Gavin Hock Tai Lim,

Tan Tock Seng Hospital, SG
X close

Lawrence Han Hwee Quek,

Tan Tock Seng Hospital, SG
X close

Uei Pua

Tan Tock Seng Hospital, SG
X close

Abstract

This article demonstrates the technique of using a coaxial guiding needle to perform combined percutaneous biopsy and microwave ablation via a single tract. From May 2019 to July 2020, 14 patients underwent combined biopsy and microwave ablation by using a coaxial guiding cannula. Tumors were in the kidney of six patients (43%), the liver of six patients (43%), and the lung in two patients (14%). The diagnostic yield of biopsy was 86% (12/14). Ablation technical success rate was 100%. In conclusion, using a coaxial guiding needle in microwave ablation and biopsy is safe and effective.

How to Cite: Wu Y-W, Chan G, Huang IKH, Kwan J, Lim GHT, Quek LHH, et al.. Combined Biopsy and Imaging-Guided Microwave Ablation by Using a Coaxial Guiding Needle. Journal of the Belgian Society of Radiology. 2021;105(1):25. DOI: http://doi.org/10.5334/jbsr.2345
111
Views
18
Downloads
  Published on 07 May 2021
 Accepted on 16 Apr 2021            Submitted on 06 Nov 2020

Introduction

In recent years, imaging-guided ablation has been used to treat tumors in different organs. Among several modalities of thermal ablation, microwave ablation (MWA) has been proven to be safe and effective in treating tumors in the liver, kidney, and lung [1, 2, 3].

Combined coaxial biopsy and ablation is such a common practice for radiofrequency ablation (RFA) [4] that an RFA device with an insulated guiding cannula is commercially available (LeVeen, Boston Scientific). At present however, there is no commercially available MWA system that provides a guiding cannula compatible with coaxial biopsy and MWA.

The aim of this article is to describe the technique of using a coaxial guiding needle to perform combined biopsy and MWA via a single path and to evaluate its feasibility.

Technique

A 17G/20 cm MWA antenna (PR 20 probe, NeuWave Medical) can be inserted coaxially through the 14G/11.6 cm guiding cannula (Figure 1A).

Figure 1 

(A) Top to bottom: inner stylet of the guiding needle, 14G/11.6 cm guiding cannula, 16G biopsy needle and 17G/20 cm MWA antenna. (B) A MWA antenna inserted through guiding cannula. Tip of the guiding cannula is indicated by the arrow. (C) CT showed a 0.8 cm metastatic nodule in the left lower lobe. (D) Microwave ablation (MWA) was performed by MWA antenna inserted through the 14G guiding cannula (arrow). (E) Post procedure CT demonstrated satisfactory penumbra covering the nodule (arrow).

During MWA procedure, a 14G/11.6 cm guiding needle was advanced into or near the tumor under ultrasound or computer tomography (CT) guidance. The inner stylet of the guiding needle was removed. Biopsy was performed with a 16G/15 cm needle coaxially through the guiding cannula. Subsequent MWA was performed with 17G/20 cm MWA antenna inserted coaxially through the guiding cannula (Figure 1B).

Patients

A total 14 consecutive patients underwent combined percutaneous biopsy and imaging-guided MWA from May 2019 to July 2020.

Technical success is defined when tumor was treated according to MWA protocol of the chosen antenna and was covered completely by ablation zone immediately after the procedure. Complications were graded based on the Cardiovascular and Interventional Radiological Society of Europe (CIRSE) classification system [5].

Patients’ characteristics and procedure details are summarized in Table 1. Overall diagnostic yield of percutaneous biopsy was 86% (12/14). Technical success was 100%. Of the 14 patients who underwent combined biopsy and MWA, two developed mild complications (14%).

Table 1

Patient characteristics, procedure details and outcome.

F female, M male, US ultrasound, CT computer tomography, HCC hepatocellular carcinoma, RCC renal cell carcinoma.


NUMBER AGE SEX ONCOLOGY HISTORY LOCATION OF TUMOR INDICATION OF BIOPSY HISTOLOGY MAXIMAL DIAMETER OF TUMOR (CM) IMAGING GUIDANCE NUMBERS OF ANTENNA TECHNICAL SUCCESS TRACT ABLATION OR EMBOLISATION COMPLICATION, GRADE

1 90 F Colonic adenocarcinoma Liver, segment III Confirm metastasis Focal regeneration 0.8 US 1 Yes No No

2 64 M No Kidney, right upper pole Confirm malignancy Papillary RCC 1.5 CT 1 Yes Tract ablation No

3 53 F Colonic adenocarcinoma Liver, segment III Confirm metastasis Non-diagnostic 2.0 US 1 Yes Tract ablation No

4 78 M No Liver, segment VI No risk factor for HCC HCC 1.6 US 1 Yes No No

5 65 M No Kidney, left lower pole Confirm malignancy Non-diagnostic 1.8 US 1 Yes Tract embolisation and ablation No

6 68 M No Liver, segment V/VI No risk factor for HCC HCC 2.5 US 1 Yes Tract ablation No

7 59 F No Liver, segment VII No risk factor for HCC Focal lobular inflammation 1.0 CT 1 Yes Tract ablation No

8 76 M No Kidney, right upper pole Confirm malignancy Papillary RCC with sarcomatoid change 4.2 CT 2 Yes Tract ablation No

9 81 M No Kidney, left lower pole Confirm malignancy Clear cell RCC 3.5 US 2 Yes No No

10 72 M No Liver, segment III No risk factor for HCC HCC 1.6 US 1 Yes No No

11 79 M No Kidney, left upper pole Confirm malignancy Papillary RCC 3.0 CT 2 Yes No No

12 78 M No Kidney, right interlobar region Confirm malignancy Clear cell RCC with sarcomatoid change 3.4 CT 2 Yes No Ablation of psoas muscle with no clinical symptom, 1

13 86 M Colonic adenocarcinoma Lung, left lower lobe Confirm metastasis Metastatic adenocarcinoma 1.0 CT 1 Yes Tract embolisation No

14 79 M Cholangiocarcinoma Lung, left upper lobe Confirm metastasis Metastatic adenocarcinoma 0.6 CT 1 Yes Tract embolisation Small pneumothorax, resolved within 48 hours, 2

Discussion

In our experience, combining biopsy and MWA through the same guiding cannula is relatively easy to perform. There are several advantages to using this technique. Firstly, biopsy is performed immediately prior to the placement of the MWA antenna. This allows the operator to better visualize and target the lesion, especially if the procedure is performed under ultrasound guidance. Using this technique, biopsy is performed by a needle with a larger diameter (16G), which may increase the diagnostic yield, as smaller needles (18G) are commonly used in tandem fashion. The biopsy diagnostic yield in our case series was 86%. Secondly, puncture-site complications (bleeding, pneumothorax, and tract seeding) are reduced, as the coaxial technique allows for performing two procedures via only one single path [4, 6]. Lastly, the coaxial guiding cannula allows for injecting a hemostatic agent to embolize the tract once ablation is completed. This is particularly useful in lung ablation, as tract embolization may reduce pneumothorax rate after the procedure [7].

One of the limitations of using this technique is the risk of performing MWA on benign lesions if the histology result is not readily available. An additional limitation is that the biopsy of a small lung nodule may cause significant alveolar hemorrhage, which may hinder the subsequent placement of MWA antenna. To overcome this, the tip of the guiding cannula can be advanced over the biopsy needle to the distal margin of the nodule. Once the biopsy needle is removed, the tip of the MWA antenna is advanced to the tip of the cannula. Maintaining the position of the MWA antenna, the guiding cannula is slowly pulled back to expose the active tip of the MWA antenna. In that case, the tip of the MWA antenna is always at the distal margin of the nodule regardless of the presence of hemorrhage.

In conclusion, using a coaxial guiding needle in MWA was safe and effective in obtaining a biopsy prior to the MWA procedure via a single path.

Competing Interests

The authors have no competing interests to declare.

References

  1. Crocetti L, de Baere T, Pereira PL, Tarantino FP. CIRSE standards of practice on thermal ablation of liver tumours. Cardiovasc Intervent Radiol. 2020; 43(7): 951–62. DOI: https://doi.org/10.1007/s00270-020-02471-z 

  2. Krokidis ME, Orsi F, Katsanos K, Helmberger T, Adam A. CIRSE guidelines on percutaneous ablation of small renal cell carcinoma. Cardiovasc Intervent Radiol. 2017; 40(2): 177–91. DOI: https://doi.org/10.1007/s00270-016-1531-y 

  3. Venturini M, Cariati M, Marra P, Masala S, Pereira PL, Carrafiello G. CIRSE standards of practice on thermal ablation of primary and secondary lung tumours. Cardiovasc Intervent Radiol. 2020; 43(5): 667–83. DOI: https://doi.org/10.1007/s00270-020-02432-6 

  4. de Baere T, Rehim MA, Teriitheau C, et al. Usefulness of guiding needles for radiofrequency ablative treatment of liver tumors. Cardiovasc Intervent Radiol. 2006; 29(4): 650–4. DOI: https://doi.org/10.1007/s00270-005-0187-9 

  5. Filippiadis DK, Binkert C, Pellerin O, Hoffmann RT, Krajina A, Pereira PL. CIRSE quality assurance document and standards for classification of complications: The CIRSE classification system. Cardiovasc Intervent Radiol. 2017; 40(8): 1141–6. DOI: https://doi.org/10.1007/s00270-017-1703-4 

  6. Wang D, Li B, Bie Z, Li Y, Li X. Synchronous core-needle biopsy and microwave ablation for highly suspicious malignant pulmonary nodule via a coaxial cannula. J Cancer Res Ther. 2019; 15(7): 1484–9. DOI: https://doi.org/10.4103/jcrt.JCRT_721_18 

  7. Tran AA, Brown SB, Rosenberg J, Hovsepian DM. Tract embolization with gelatin sponge slurry for prevention of pneumothorax after percutaneous computed tomography-guided lung biopsy. Cardiovasc Intervent Radiol. 2014; 37(6): 1546–53. DOI: https://doi.org/10.1007/s00270-013-0823-8 